![]() |
|||||
![]() |
|||||
![]() |
|||||
Start Europa - katalog-czesci-[DE], rolnicze, •• MASZYNY rolnicze - instrukcje serwisowe itp Ergoline EXCELLENCE 700, solarium Ergoline 500,600,700,800,900 - instrukcja po polsku Ergoline Classic E 600 Manual, solarium Ergoline 500,600,700,800,900 - instrukcja po polsku Escorregando Tango Brasileiro - Piano, NUUUTY, INSTRUMENT, Tango Europa swiat spr 1B INSTRUKCJA, Szkoła, Liceum, His - Historia i Społeczetwo, Sprawdziany Ethernet Gateway, Instrukcje, EATON MOELLER Every Breath You Take Notes, nauka gry na instrumencie ERC, INSTRUKCJE SERVISOWE BUDERUS Ethernet%20Crossover, Instrukcje, PDF Ethernet%20Straight, Instrukcje, PDF |
EWDR 983 985, Instrukcje, ELIWELL[ Pobierz całość w formacie PDF ]EWDR 983-985 /CS (LX) electronic controllers for ventilated refrigeration units cod. 9IS43068 rel. 1/05 GB USER INTERFACE The user interface has a 6 LED display to indicate alarm status and five buttons for controlling instrument status and pro- gramming. SIGNAL LED The status of the external devices, func- tions and controllers is described by the device LEDs. The set point setting folder is also closed when the time-out elapses. Alarm on If an alarm condition exists when the Machine Status menu is accessed the “ALfolder label appears. DISPLAY This is used to display the inputs, the set point, the parameters and related values, alarms, functions and the status of the device. Real Time Clock (/C models) By pressing the “set” button when the “rtc” label appears, the label d00 (days) is displayed. Use the “UP” and “DOWN” buttons to set days. If you do not use the buttons for over 2 seconds or if you press “set” you switch to the hours (h00) and minutes (‘00) folders: use the “UP” and “DOWN” buttons to set the hours and minutes respectively. If you do not use the keyboard for over 15 seconds (time-out) or if you press the “fnc” but- ton once, the last value shown on the display is confirmed and you go back to the previous screen mask. NOTE: Always use the “set” button to confirm the hours/minutes/days set- ting. NOTE2: We recommend considering the first day d00 as SUNDAY. DESCRIPTION OF MENU Access to both menus is controlled by the ‘set’ button. If it is pressed and immediately released, the ‘machine sta- tus menu‘ is displayed. Hold the same button down for 5 seconds to access the ‘parameter programming menu‘. When one of the two menus has been accessed, you can navigate between the level 1 folders using the ‘UP’ and ‘DOWN’ buttons. The folders are opened by pressing the ‘set’ button once. You can now scroll through the contents of each folder, modify it or use its functions. You can exit each level of both menus in three ways: using the ‘fnc’ button, if a new value is confirmed by pressing the ‘set’ button or, when the time-out has elapsed (15 seconds inactivity on the device). BUTTONS AND LEDS UP Increase in value of parameter Menu scrolling and activation of defrost function (Parameter programmable H31) DOWN Decrease in value of parameter Menu scrolling and activation of associated function (Parameter programmable H32) set Access to different levels of menu Alarm display, set point, probes and rtc Access to programming of parameters PARAMETER PROGRAMMING MENU Access the menu by pressing the ‘set’ button for at least 5 seconds. The menu structure enables all parameter folders to be divided into two levels. All the level 1 folders can be accessed by enter- ing the password ‘PA1’. Scroll down the level 1 folders using the ‘UP’ and ‘DOWN’ buttons. Press and release the ‘set’ button next to the selected label to access the parameters. Scroll through the labels in the folder using the ‘UP’ and ‘DOWN’ buttons, press ‘set’ to display the current value of the selected parameter, use the ‘UP’ and ‘DOWN’ buttons and set the required value by pressing ‘set’. To access the level 2 folders in the ‘Cnf’ folder, select the ‘PA2’ label, enter the password ‘PA2’ and confirm with the ‘set’ button. All the parameters that cannot be changed at level 1 are in this level. NOTE: Level 1 parameters will only be displayed if you quit the ‘parameter programming menu’ and repeat the steps for manipulation of level 1 folders. The steps to follow for manipulation of level 2 parameters are the same as those described for the level 1 structure. NOTE: It is strongly recommended that the instrument is switched off and on again each time parameter MACHINE STATUS MENU The ‘machine status menu’ contains the folders and basic information on the device: -AL: alarm folder -SEt: Set point setting folder -rtc: Real Time Clock folder(if present) -Pb1: ‘probe 1 value’ folder -Pb2: ‘probe 2 value’ folder -Pb3: ‘probe 3 value’ folder fnc Exit current level of menu Confirmation of value of parameter (Parameter programmable H33) aux Activation of associated function using parameter H34 LED Description Indication If no alarms are present, the “SEt” label is displayed. From here, you can scroll down the other menu items using the UP’ and ‘DOWN’ buttons. Each folder can be accessed by pressing the ‘set’ button once. Values are modified using the ‘UP’ and ‘DOWN’ buttons and the ‘set’ button that confirms the selected value and takes you back to the higher level. Setting the set point Access the ‘machine status menu’. If no alarms are present, the “SEt” label is dis- played. By pressing and immediately releasing the ‘set’ button, the set point value can be set using the UP’ and ‘DOWN’ buttons. Press and release the ‘set’ button again or press the ‘fnc’ but- ton to go back to the main menu level. Reduced set point Lights up when inserted LED set point is displayed, blinking when reduced set point is inserted LED ‘on’ with compressor LED on. Blinking for delay, protection or activation blocked defrost LED ‘on’ during defrosting Blinking when activated manually or by digital input fan LED permanently on when fans are operating alarm LED ‘on’ if alarm is present. Blinking for silenced alarm auxiliary LED ‘on’ for active output decimal point ‘on’ to indicate of voltage when on stand-by and display ‘off’ configuration is changed in order to prevent malfunctioning of the config- uration and/or ongoing timings. Televis remote control systems can be connected using the RS 485 serial input. To configure the instrument to do this, you need to access the folder with the “Add” label and use the “dEA” and “FAA” parameters. Parameter H06 activates the light button and enables the relay when the door is open even when the instrument is pow- ered but is ‘off’. The dedicated button always deactivates the light relay even if the digital input is on or during the dLt delay only if the parameter OFL=Y. PASSWORD The passwords “PA1and “PA2” are used to access level 1 and level 2 parameters. To change them and assign them the desired value, access the ‘parameter pro- gramming menu’ in the “diS” label folder. ADVANCED FUNCTIONS DIAGNOSTICS PROBE ALARMS When one of the probes is outside the nominal operating range or the probe is open or has shorted, an alarm is gener- ated if at least one of these conditions persists for at least 10 seconds. If at least one of these 3 alarms is sig- nalled the alarm LED and relay are enabled. The password is requested: - PA1 when entering the ‘parameter pro- gramming menu’; - PA2 in the “Cnf” folder containing the level 1 parameters. DOOR SWITCH INPUT This is a clean contact digital input with programmable polarity. The door switch input functions are controlled by the values of the following parameters: Par Description dOd Digital input switches off loads dAd USING THE COPY CARD The Copy card function can be used to upload and download parameter maps of one or more of the same type of instrument. The accessory is connected to the device using the special TTL serial port on the instrument. The Copy card can be used for the following functions: Upload (UL) The upload function reads the instru- ment parameters and writes them onto the Copy card Download(dL) The download function writes the para- meter map present at that time on the copy card onto the instrument memory. NOTE: Upload = instrument—-> Copy card Download = Copy card —-> instrument The operations are performed by access- ing the folder with the ‘FPr’ label and selecting the ‘UL’, ’dL’ or ‘Fr’ com- mands. The operation is confirmed by pressing the ‘set’ button. If the operation is successful, a “y” is displayed whereas if it is unsuccessful an “n” will be dis- played. Formatting the copy card This function is necessary when using the copy card for uploading for the first time or when using the copy card on device models that are not compatible. NOTE: formatting deletes all the data on the Copy card and cannot be undone. Download from reset Once the copy card has been connected the instrument is switched on. When the lamp test is over, one of the following two labels will be displayed for 5 sec- onds: - dLY if the operation is successful - label DLn if operation fails The display will then go into default position (probe or set point). NOTE: after downloading the instru- ment will begin to work with the new parameter map that has just been downloaded. REMOTE CONTROL SYSTEMS (/CS LX models only) D.I. activation delay OAO Alarm signal delay after disdisabling the digital input (door closed) tdO Time out door open. Time out signal after activation of D.I. (door open) Signal Description E1 Faulty room probe E2 Faulty defrost probe H11 Digital output configurability/polarity 1 E3 Faulty display probe H21...H25 Digital output configurability 1...5 When activated code E1 appears on the display. If several alarms are activated at the same time, they are displayed alter- nately for 2 seconds each. An error condition in the room probe leads to: - E1 code appears on display - activation of compressor as indicated by Ont and OFt - deactivation of maximum and mini- mum alarm controller. An error condi- tion in the defrost probe leads to: - E2 code appearing on display - end of defrost due to a time-out Parameter H11 is used to configure the door switch input with values between -8 and +8. Positive and negative values are present in order to select the polari- ty assigned to the input: NOTE: the ‘-’ sign indicates that the input is activated when the contact is closed The ‘+’ sign indicates that the input is activated when the contact is open. LIGHT CONTROLLER Controls the light relay. The function can be started in two different ways. If you press the light button the light relay is switched on if it was off and is switched off if it was on. The light status is recorded by the device as soon as the button is pressed so that when power is restored after a black-out the device can continue to operate in the same way as before the power failure. The following parameters control how the light works: Par Description H06 Button/aux input/door switch light active when instrument off dSd Enabling light relay by door switch dLt Delay Light Relay deactivation delay OFL Light switch always disables light relay OAO Alarm signal delay after disabling the digital input (door closed) tdO Time out door open. Time out signal after activation of D.I. (door open) H11 Digital output configurability/polarity 1 H21...H25 Digital output configurability 1...5 The configuration of these parameters controls the status of the light relay by using the digital input as well as the nor- mal button. The parameter dSd automat- ically switches on the light relay when the digital input is enabled and switches it off when the digital input is disabled following the delay set by parameter dLt. EWDR 983-985 /CS (LX) 2/11 EXTERNAL ALARM It is set if the digital input is enabled with the delay specified by parameter dAd and remains enabled until the next time the digital input is deactivated. The alarm consists of an alarm LED that remains on, activation of a buzzer and deactivation of all the device loads (if specified by the EAL parameter). If an external alarm is present the EA label is displayed in the AL folder. Serial: -TTL for Copy Card -RS485 for connection to Televis System (for /CS LX models only) Digital outputs: EWDR 983 (/CS LX): 2 digital outputs on relay:8A 1/2hp 250V SPDT; 16A 1hp 250V SPDT; EWDR 985 (/CS LX): 4 digital outputs on relays: 8A 1/2hp 250V SPST; 8A 1/2hp 250V SPDT; 16A 1hp 250V SPST; 8A 1/2hp 250V SPST. Configurable 12Vc/24mA output Resolution: 1 or 0.1°C. Accuracy: better than 0.5% of bottom scale +1 digit Consumption: 5VA Power supply: 230V MINIMUM AND MAXIMUM TEMPERA- TURE ALARM The alarm is regulated on the room probe. The temperature limits are defined by parameters HAL and LAL. The limits refer to the set point if para- meter Att=1 and are absolute if Att=0. Att=0 absolute Maximum temperature alarm room probe temperature ≥ HAL Minimum temperature alarm room probe temperature ≤ LAL Signal Description EA External alarm Att=1 relative Maximum temperature alarm room probe temperature ≥ Set + HAL It is reset the next time the digital input is deactivated and the buzzer can be manually silenced. Minimum temperature alarm room probe temperature ≤ Set + LAL NOTE: If the alarms are relative, the parameter HA1 is set to positive val- ues and LA1 to negative values. OPEN DOOR ALARM This alarm is generated when the door is open for longer than the value specified in parameter tdO. door open alarm is signalled in the alarm folder (AL) and the LED and alarm relay are switched on. The OPd label is also displayed. a CLOCK BATTERY ALARM This alarm is activated in the following situations: - when the instrument is switched on for the first time and the clock is set - when the instrument is switched on again after a power off lasting more than 24/32 hours. ELECTRICAL CONNECTIONS Warning! Always switch off machine before working on electrical connec- tions. The instrument has screw terminal blocks for connecting cables with a maximum diameter of 2.5 mm 2 (only one conductor per terminal block for power connec- tions): for terminal capacity, see the label on the instrument. The relay outputs are voltage free. Do not exceed the maximum current allowed. For higher loads, use a suitable contactor. Make sure that the power voltage com- plies with the device voltage. Probes have no connection polarity and can be extended using an ordinary bipo- lar cable (note that if probes are extend- ed this affects the electromagnetic com- patibility (EMC) of the instrument: special care must be used when wiring). The RS-485 serial and TTL serial cables must be kept separate from the power cables. Signal Description OPd Open door alarm The relay can be deactivated by pressing the silencing button. The LED continues to flash intermittently and the OPd label remains in the alarm menu until the door is closed. Signal Description E10 Loss of hours with buffer battery The alarm will stop as soon as the day/hour/minutes are set using the set parameters. Resetting is automatic and there is no need to restart the instrument. MECHANICAL ASSEMBLY The EWDR 983-985 LX is designed to be wall-mounted (removable brackets) or mounted on DIN rail (Omega 3). The unit operates correctly with an ambient temperature range of between -5 and 55 °C. Units must not be installed in excessively humid and/or dirty loca- tions. ALARM WITH THRESHOLD REFERRING TO PROBE 3 An alarm is associated with probe 3 that refers to a threshold that is reset at a specified differential. By setting parameter PbA=3 probe 3 will signal a high or low temperature alarm for exceeding the set value. TECHNICAL DATA Casing: plastic 4 Din modules Dimensions: front 70x85 mm, depth 61mm. Mounting: on DIN rail (Omega 3) or wall-mounted. Connections: on screw terminal block for ≤ 2.5mm 2 conductors (only one conductor per terminal block for power connections) Operating temperature: -5...55°C. Storage temperature: -30...85°C. Operating and storage ambient humidity: 10...90 % RH (non-condensing). Display: 3 digits + sign Analogue inputs: 3 NTC/PTC inputs (parameter configurable) Digital inputs: 2 voltage-free parameter- configurable digital inputs. Signal Description AH3 High temperature alarm AL3 Low temperature alarm The alarm is handled in the same way as the other temperature alarms and stan- dard signal delays are used. DEFROST ALARM If defrosting does not reach the end of cycle temperature but is terminated by a time out, the alarm controller is immedi- ately activated. An alarm LED remains on and the alarm is recorded in the Ad2 label alarm menu. Signal Description Ad2 End of defrost due to time out The alarm is automatically reset as soon as the next defrost cycle starts. The alarm LED can only be manually switched off with a silencing procedure. EWDR 983-985 /CS (LX) 3/11 PAR. DESCRIPTION RANGE DEFAULT VALUE LEVEL U.M. SEt Set point with range falling between the minimum LSE set point and the maximum HSE set point. The value of the set point is in the machine status menu LSE...HSE 0.0 °C/°F diF When the compressor stops when it reaches the set point value, it restarts at a value corresponding to the set point plus the value of the differential It must not be 0 0.1...30.0 2.0 1-2 °C/°F HSE Maximum set point value LSE...302 50.0 1-2 °C/°F LSE Minimum set point value -58.0...HSE -50.0 1-2 °C/°F OSP Offset point. Value to be added to set point if reduced set point is activated Economy function). -30.0...30.0 0 2 °C/°F Cit Minimum compressor ON time. Minimum time for activating a compressor before deactivation Not active if=0 0...250 0 2 min CAt Maximum compressor ON time. Maximum time for activating a compressor before deactivation Not active if=0 0...250 0 2 min Ont (1) Compressor activation time if probe is faulty. If set to 1 with OFt=0 the compressor always remains on whereas if OFt>0 it operates in duty cycle mode (see Duty Cycle diagram) 0...250 0 1-2 min OFt (1) Compressor shut-down time if probe is faulty. If set to 1 with Ont=0 the compressor always remains off whereas if OFt>0 it operates in duty cycle mode (see Duty Cycle diagram) 0...250 1 1-2 min dOn Delay in activating compressor relay after start-up 0...250 0 1-2 sec dOF Delay after shut-down; between compressor relay shut-down and subsequent start-up the specified time must elapse. 0...250 0 1-2 min dbi Delay between switch-ons; the specified time must elapse between two subsequent switch-ons 0...250 0 1-2 min OdO Delay in activating outputs after start-up of instru- ment or after a power failure. Not active if=0 0...250 0 1-2 min dty Type of defrost. 0 = electrical defrosting; 1 = cycle reversing defrosting (hot gas); 2 = Free mode defrosting (compressor disabled). 0/1/2 0 1-2 flag dit Period of time elapsing between the start of two defrosts 0=function disabled 0...250 6 1-2 hours dt1 Unit of measurement for defrost times (par. dit) 0=”dit” expressed in hours 1=”dit” expressed in minutes 2=”dit” expressed in seconds 0/1/2 0 2 flag dt2 Unit of measurement for duration of defrosting (dEt parameter) 0= “dEt” parameter expressed in hours 1= “dEt” parameter expressed in minutes 2= “dEt” parameter expressed in seconds 0/1/2 1 2 flag dCt Selection of defrosting time count mode. 0=compressor operating hours DIGIFROST® method). Defrosting active only if compressor is on. 1 = equipment operating hours; defrost counting is always active when the machine is on 2=compressor stop Each time the compressor stops a defrosting cycle according to par. dtY 3=con RTC. Defrosting at times set by dE1...dE8, F1...F8 parameters. 0/1/2/3 1 1-2 flag dOH Delay between start of first defrosting operation and start-up of instrument. 0...59 0 1-2 min dEt Defrosting time-out; determines maximum dura- tion of defrosting. 1...250 30 1-2 min dE2 Defrosting time-out on second evaporator; deter- mines maximum duration of defrosting on 2nd evaporator 1...250 30 2 min NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels EWDR 983-985 /CS (LX) 4/11 PAR. DESCRIPTION RANGE DEFAULT VALUE* LEVEL** U.M. dSt End of defrost temperature (determined by evapo- rator probe) -50.0...150 8.0 1-2 °C/°F dS2 End of defrost temperature (determined from probe on 2nd evaporator) -50.0...150 8.0 2 °C/°F dPO Determines when instrument starts up if the defrosting cycle must be activated (if the tempera- ture on the evaporator allows this) y=defrosting activated at start-up n=defrosting not activated at start-up n/y n 1-2 flag tcd minimum time for each compressor state before defrosting “Ontime if >0; “Offtime if >0 -31...31 0 2 min Cod Compressor “Off” time before defrost cycle. The compressor is not turned on if a defrost cycle is expected in the time indicated by the parameter. 0=Function excluded 0...60 0 2 min “dd (2) dE1...dE8; daily defrost start time 0...23/0...59 24 1-2 hours/min “Fd (2) F1...F8 festive defrost start time 0...23/0...59 24 1-2 hours/min FPt Determines if “FSt” and “Fot” are expressed as absolute values or in relation to set point 0=absolute value; 1=value related to set point 0/1 0 2 flag FSt Fan stop temperature. Temperature limit that, if exceeded by the value read by the evaporator probe, stops the fans. -50.0...150.0 2.0 1-2 °C/°F Fot Fan start temperature. If the temperature read by the evaporator probe is lower than the set value the fans remain off. -50.0...150.0 -50.0 2 °C/°F FAd Fan activation intervention differential. (see “FSt”, “Fot”) 1.0...50.0 2.0 1-2 °C/°F Fdt Delay before fan activation after defrosting 0...250 0 1-2 min dt Dripping time min 0...250 0 1-2 dFd Disables evaporator fans. y=fans disabled n=fans enabled y/n y 1-2 flag FCO Disables fans with compressor off (Off) y = fans active (with thermostat; in response to value read by defrost probe, see “FSt” parameter); n = fans off; dc = duty cycle (using parameters “Fon” and “FoF”) n/y/dc y 1-2 flag Fod Enables fan stop with door open and fan re-start when door is closed (if fans were on). n=fans stop; y=fans unchanged n/y n 2 flag FdC Fan shut-down delay after compressor stop 0=function excluded 0...99 0 2 min Fon Fan start-up time in Duty Cycle mode; valid for FCO=dc and H42=1 (presence of probe 2) 0...99 0 2 min FoF Fan shut-down time in Duty Cycle mode; valid for FCO=dc and H42=1 (presence of probe 2) 0...99 0 2 min Att Determines if “LAL” and “HAL” are expressed as absolute values or as a differential related to the set point 0=absolute value 1=value related to set point 0/1 0 2 flag AFd Alarm differential 1.0...50.0 2.0 1-2 °C/°F HAL (4) Maximum alarm. Temperature limit (whose absolute or relative value status is regulated by “Att”) above which the alarm is activated. LAL...150.0 50.0 1-2 °C/°F LAL (4) Minimum alarm. Temperature limit (whose absolute or relative value status is regulated by “Att”) below which the alarm is activated. -50.0...HAL -50.0 1-2 °C/°F PAO (5) Alarm exclusion time after start-up of instrument following a power failure 0...10 0 1-2 hours dAO Alarm exclusion time after defrosting 0...999 0 1-2 min NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels EWDR 983-985 /CS (LX) 5/11 [ Pobierz całość w formacie PDF ] |
||||
![]() |
|||||
Wszelkie Prawa Zastrzeżone! Oto smutna prawda: cierpienie uszlachetnia. Design by SZABLONY.maniak.pl. |
![]() |
||||
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |