Start
Europa - katalog-czesci-[DE], rolnicze, •• MASZYNY rolnicze - instrukcje serwisowe itp
Ergoline EXCELLENCE 700, solarium Ergoline 500,600,700,800,900 - instrukcja po polsku
Ergoline Classic E 600 Manual, solarium Ergoline 500,600,700,800,900 - instrukcja po polsku
Escorregando Tango Brasileiro - Piano, NUUUTY, INSTRUMENT, Tango
Europa swiat spr 1B INSTRUKCJA, Szkoła, Liceum, His - Historia i Społeczetwo, Sprawdziany
Ethernet Gateway, Instrukcje, EATON MOELLER
Every Breath You Take Notes, nauka gry na instrumencie
ERC, INSTRUKCJE SERVISOWE BUDERUS
Ethernet%20Crossover, Instrukcje, PDF
Ethernet%20Straight, Instrukcje, PDF
  • zanotowane.pl
  • doc.pisz.pl
  • pdf.pisz.pl
  • ets2.xlx.pl

  • EWDR 983 985, Instrukcje, ELIWELL

    [ Pobierz całość w formacie PDF ]
    EWDR 983-985 /CS (LX)
    electronic controllers for ventilated refrigeration units
    cod. 9IS43068
    rel. 1/05
    GB
    USER INTERFACE
    The user interface has a 6 LED display to
    indicate alarm status and five buttons for
    controlling instrument status and pro-
    gramming.
    SIGNAL LED
    The status of the external devices, func-
    tions and controllers is described by the
    device LEDs.
    The set point setting folder is also closed
    when the time-out elapses.
    Alarm on
    If an alarm condition exists when the
    Machine Status menu is accessed the
    “ALfolder label appears.
    DISPLAY
    This is used to display the inputs, the set
    point, the parameters and related values,
    alarms, functions and the status of the
    device.
    Real Time Clock (/C models)
    By pressing the “set” button when the
    “rtc” label appears, the label d00 (days)
    is displayed. Use the “UP” and “DOWN”
    buttons to set days. If you do not use
    the buttons for over 2 seconds or if you
    press “set” you switch to the hours (h00)
    and minutes (‘00) folders: use the “UP”
    and “DOWN” buttons to set the hours
    and minutes respectively. If you do not
    use the keyboard for over 15 seconds
    (time-out) or if you press the “fnc” but-
    ton once, the last value shown on the
    display is confirmed and you go back to
    the previous screen mask.
    NOTE: Always use the “set” button to
    confirm the hours/minutes/days set-
    ting.
    NOTE2: We recommend considering
    the first day d00 as SUNDAY.
    DESCRIPTION OF MENU
    Access to both menus is controlled by
    the ‘set’ button. If it is pressed and
    immediately released, the ‘machine sta-
    tus menu‘ is displayed. Hold the same
    button down for 5 seconds to access the
    ‘parameter programming menu‘.
    When one of the two menus has been
    accessed, you can navigate between the
    level 1 folders using the ‘UP’ and ‘DOWN’
    buttons. The folders are opened by
    pressing the ‘set’ button once. You can
    now scroll through the contents of each
    folder, modify it or use its functions.
    You can exit each level of both menus in
    three ways: using the ‘fnc’ button, if a
    new value is confirmed by pressing the
    ‘set’ button or, when the time-out has
    elapsed (15 seconds inactivity on the
    device).
    BUTTONS AND LEDS
    UP
    Increase in value of
    parameter Menu scrolling
    and activation of defrost
    function
    (Parameter programmable
    H31)
    DOWN
    Decrease in value of
    parameter
    Menu scrolling and activation
    of associated function
    (Parameter programmable
    H32)
    set
    Access to different levels of
    menu Alarm display, set
    point, probes and rtc
    Access to programming of
    parameters
    PARAMETER PROGRAMMING MENU
    Access the menu by pressing the ‘set’
    button for at least 5 seconds. The menu
    structure enables all parameter folders
    to be divided into two levels. All the
    level 1 folders can be accessed by enter-
    ing the password ‘PA1’.
    Scroll down the level 1 folders using the
    ‘UP’ and ‘DOWN’ buttons. Press and
    release the ‘set’ button next to the
    selected label to access the parameters.
    Scroll through the labels in the folder
    using the ‘UP’ and ‘DOWN’ buttons,
    press ‘set’ to display the current value of
    the selected parameter, use the ‘UP’ and
    ‘DOWN’ buttons and set the required
    value by pressing ‘set’. To access the
    level 2 folders in the ‘Cnf’ folder, select
    the ‘PA2’ label, enter the password ‘PA2’
    and confirm with the ‘set’ button. All the
    parameters that cannot be changed at
    level 1 are in this level.
    NOTE: Level 1 parameters will only be
    displayed if you quit the ‘parameter
    programming menu’ and repeat the
    steps for manipulation of level 1 folders.
    The steps to follow for manipulation of
    level 2 parameters are the same as those
    described for the level 1 structure.
    NOTE: It is strongly recommended
    that the instrument is switched off
    and on again each time parameter
    MACHINE STATUS MENU
    The ‘machine status menu’ contains the
    folders and basic information on the
    device:
    -AL: alarm folder
    -SEt: Set point setting folder
    -rtc: Real Time Clock folder(if present)
    -Pb1: ‘probe 1 value’ folder
    -Pb2: ‘probe 2 value’ folder
    -Pb3: ‘probe 3 value’ folder
    fnc
    Exit current level of
    menu Confirmation of value of
    parameter
    (Parameter programmable
    H33)
    aux
    Activation of associated
    function using parameter
    H34
    LED
    Description
    Indication
    If no alarms are present, the “SEt” label
    is displayed. From here, you can scroll
    down the other menu items using the
    UP’ and ‘DOWN’ buttons. Each folder
    can be accessed by pressing the ‘set’
    button once.
    Values are modified using the ‘UP’ and
    ‘DOWN’ buttons and the ‘set’ button
    that confirms the selected value and
    takes you back to the higher level.
    Setting the set point
    Access the ‘machine status menu’. If no
    alarms are present, the “SEt” label is dis-
    played. By pressing and immediately
    releasing the ‘set’ button, the set point
    value can be set using the UP’ and
    ‘DOWN’ buttons. Press and release the
    ‘set’ button again or press the ‘fnc’ but-
    ton to go back to the main menu level.
    Reduced set point
    Lights up when
    inserted LED
    set point is
    displayed, blinking
    when reduced set
    point is inserted
    LED
    ‘on’ with compressor
    LED on. Blinking
    for delay, protection
    or activation blocked
    defrost LED
    ‘on’ during defrosting
    Blinking when
    activated manually
    or by digital input
    fan LED
    permanently on when
    fans are operating
    alarm LED
    ‘on’ if alarm is present.
    Blinking for silenced
    alarm
    auxiliary LED
    ‘on’ for active output
    decimal point
    ‘on’ to indicate of
    voltage when on
    stand-by and display
    ‘off’
     configuration is changed in order to
    prevent malfunctioning of the config-
    uration and/or ongoing timings.
    Televis remote control systems can be
    connected using the RS 485 serial input.
    To configure the instrument to do this,
    you need to access the folder with the
    “Add” label and use the “dEA” and “FAA”
    parameters.
    Parameter H06 activates the light button
    and enables the relay when the door is
    open even when the instrument is pow-
    ered but is ‘off’. The dedicated button
    always deactivates the light relay even if
    the digital input is on or during the dLt
    delay only if the parameter OFL=Y.
    PASSWORD
    The passwords “PA1and “PA2” are used
    to access level 1 and level 2 parameters.
    To change them and assign them the
    desired value, access the ‘parameter pro-
    gramming menu’ in the “diS” label folder.
    ADVANCED FUNCTIONS
    DIAGNOSTICS
    PROBE ALARMS
    When one of the probes is outside the
    nominal operating range or the probe is
    open or has shorted, an alarm is gener-
    ated if at least one of these conditions
    persists for at least 10 seconds.
    If at least one of these 3 alarms is sig-
    nalled the alarm LED and relay are
    enabled.
    The password is requested:
    - PA1 when entering the ‘parameter pro-
    gramming menu’;
    - PA2 in the “Cnf” folder containing the
    level 1 parameters.
    DOOR SWITCH INPUT
    This is a clean contact digital input with
    programmable polarity. The door switch
    input functions are controlled by the
    values of the following parameters:
    Par Description
    dOd Digital input switches off loads
    dAd
    USING THE COPY CARD
    The Copy card function can be used to
    upload and download parameter maps
    of one or more of the same type of
    instrument. The accessory is connected
    to the device using the special TTL serial
    port on the instrument. The Copy card
    can be used for the following functions:
    Upload (UL)
    The upload function reads the instru-
    ment parameters and writes them onto
    the Copy card
    Download(dL)
    The download function writes the para-
    meter map present at that time on the
    copy card onto the instrument memory.
    NOTE:
    Upload = instrument—-> Copy card
    Download = Copy card —-> instrument
    The operations are performed by access-
    ing the folder with the ‘FPr’ label and
    selecting the ‘UL’, ’dL’ or ‘Fr’ com-
    mands. The operation is confirmed by
    pressing the ‘set’ button. If the operation
    is successful, a “y” is displayed whereas if
    it is unsuccessful an “n” will be dis-
    played.
    Formatting the copy card
    This function is necessary when using the
    copy card for uploading for the first
    time or when using the copy card on
    device models that are not compatible.
    NOTE: formatting deletes all the data
    on the Copy card and cannot be
    undone. Download from reset
    Once the copy card has been connected
    the instrument is switched on. When the
    lamp test is over, one of the following
    two labels will be displayed for 5 sec-
    onds:
    - dLY if the operation is successful
    - label DLn if operation fails The display
    will then go into default position (probe
    or set point).
    NOTE: after downloading the instru-
    ment will begin to work with the new
    parameter map that has just been
    downloaded.
    REMOTE CONTROL SYSTEMS
    (/CS LX models only)
    D.I. activation delay
    OAO
    Alarm signal delay after disdisabling the
    digital input (door closed)
    tdO
    Time out door open. Time out
    signal after activation of D.I.
    (door open)
    Signal
    Description
    E1
    Faulty room probe
    E2
    Faulty defrost probe
    H11
    Digital output configurability/polarity 1
    E3
    Faulty display probe
    H21...H25
    Digital output configurability 1...5
    When activated code E1 appears on the
    display. If several alarms are activated at
    the same time, they are displayed alter-
    nately for 2 seconds each.
    An error condition in the room probe
    leads to:
    - E1 code appears on display
    - activation of compressor as indicated
    by Ont and OFt
    - deactivation of maximum and mini-
    mum alarm controller. An error condi-
    tion in the defrost probe leads to:
    - E2 code appearing on display
    - end of defrost due to a time-out
    Parameter H11 is used to configure the
    door switch input with values between
    -8 and +8. Positive and negative values
    are present in order to select the polari-
    ty assigned to the input:
    NOTE:
    the ‘-’ sign indicates that the input is
    activated when the contact is closed
    The ‘+’ sign indicates that the input is
    activated when the contact is open.
    LIGHT CONTROLLER
    Controls the light relay. The function can
    be started in two different ways. If you
    press the light button the light relay is
    switched on if it was off and is switched
    off if it was on. The light status is
    recorded by the device as soon as the
    button is pressed so that when power is
    restored after a black-out the device can
    continue to operate in the same way as
    before the power failure. The following
    parameters control how the light works:
    Par
    Description
    H06
    Button/aux input/door switch light active
    when instrument off
    dSd
    Enabling light relay by door switch
    dLt
    Delay Light Relay deactivation delay
    OFL
    Light switch always disables light relay
    OAO
    Alarm signal delay after disabling the
    digital input (door closed)
    tdO
    Time out door open. Time out
    signal after activation of D.I.
    (door open)
    H11
    Digital output configurability/polarity 1
    H21...H25
    Digital output configurability 1...5
    The configuration of these parameters
    controls the status of the light relay by
    using the digital input as well as the nor-
    mal button. The parameter dSd automat-
    ically switches on the light relay when
    the digital input is enabled and switches
    it off when the digital input is disabled
    following the delay set by parameter dLt.
    EWDR 983-985 /CS (LX)
    2/11
     EXTERNAL ALARM
    It is set if the digital input is enabled
    with the delay specified by parameter
    dAd and remains enabled until the next
    time the digital input is deactivated.
    The alarm consists of an alarm LED that
    remains on, activation of a buzzer and
    deactivation of all the device loads (if
    specified by the EAL parameter).
    If an external alarm is present the EA
    label is displayed in the AL folder.
    Serial:
    -TTL for Copy Card
    -RS485 for connection to Televis
    System
    (for /CS LX models only)
    Digital outputs:
    EWDR 983 (/CS LX):
    2 digital outputs on relay:8A 1/2hp 250V
    SPDT; 16A 1hp 250V SPDT;
    EWDR 985 (/CS LX):
    4 digital outputs on relays: 8A 1/2hp
    250V SPST; 8A 1/2hp 250V SPDT; 16A
    1hp 250V SPST; 8A 1/2hp 250V SPST.
    Configurable 12Vc/24mA output
    Resolution: 1 or 0.1°C.
    Accuracy: better than 0.5% of bottom
    scale +1 digit
    Consumption: 5VA
    Power supply: 230V
    MINIMUM AND MAXIMUM TEMPERA-
    TURE ALARM
    The alarm is regulated on the room
    probe. The temperature limits are
    defined by parameters HAL and LAL.
    The limits refer to the set point if para-
    meter Att=1 and are absolute if Att=0.
    Att=0
    absolute
    Maximum temperature alarm
    room probe temperature

    HAL
    Minimum temperature alarm
    room probe temperature

    LAL
    Signal
    Description
    EA
    External alarm
    Att=1
    relative
    Maximum temperature alarm
    room probe temperature

    Set + HAL
    It is reset the next time the digital input
    is deactivated and the buzzer can be
    manually silenced.
    Minimum temperature alarm
    room probe temperature

    Set + LAL
    NOTE: If the alarms are relative, the
    parameter HA1 is set to positive val-
    ues and LA1 to negative values.
    OPEN DOOR ALARM
    This alarm is generated when the door is
    open for longer than the value specified
    in parameter tdO.
    door open alarm is signalled in the
    alarm folder (AL) and the LED and alarm
    relay are switched on.
    The OPd label is also displayed.
    a
    CLOCK BATTERY ALARM
    This alarm is activated in the following
    situations:
    - when the instrument is switched on for
    the first time and the clock is set
    - when the instrument is switched on
    again after a power off lasting more than
    24/32 hours.
    ELECTRICAL
    CONNECTIONS
    Warning! Always switch off machine
    before working on electrical connec-
    tions.
    The instrument has screw terminal blocks
    for connecting cables with a maximum
    diameter of 2.5 mm
    2
    (only one conductor
    per terminal block for power connec-
    tions): for terminal capacity, see the label
    on the instrument.
    The relay outputs are voltage free.
    Do not exceed the maximum current
    allowed. For higher loads, use a suitable
    contactor.
    Make sure that the power voltage com-
    plies with the device voltage.
    Probes have no connection polarity and
    can be extended using an ordinary bipo-
    lar cable (note that if probes are extend-
    ed this affects the electromagnetic com-
    patibility (EMC) of the instrument: special
    care must be used when wiring).
    The RS-485 serial and TTL serial cables
    must be kept separate from the power
    cables.
    Signal
    Description
    OPd
    Open door alarm
    The relay can be deactivated by pressing
    the silencing button. The LED continues
    to flash intermittently and the OPd label
    remains in the alarm menu until the
    door is closed.
    Signal
    Description
    E10
    Loss of hours with buffer
    battery
    The alarm will stop as soon as the
    day/hour/minutes are set using the set
    parameters.
    Resetting is automatic and there is no
    need to restart the instrument.
    MECHANICAL ASSEMBLY
    The EWDR 983-985 LX is designed to be
    wall-mounted (removable brackets) or
    mounted on DIN rail (Omega 3).
    The unit operates correctly with an
    ambient temperature range of between
    -5 and 55 °C. Units must not be installed
    in excessively humid and/or dirty loca-
    tions.
    ALARM WITH THRESHOLD REFERRING
    TO PROBE 3
    An alarm is associated with probe 3 that
    refers to a threshold that is reset at a
    specified differential.
    By setting parameter PbA=3 probe 3 will
    signal a high or low temperature alarm
    for exceeding the set value.
    TECHNICAL DATA
    Casing: plastic 4 Din modules
    Dimensions: front 70x85 mm, depth
    61mm.
    Mounting: on DIN rail (Omega 3) or
    wall-mounted.
    Connections: on screw terminal block for

    2.5mm
    2
    conductors
    (only one conductor per terminal block
    for power connections)
    Operating temperature: -5...55°C.
    Storage temperature: -30...85°C.
    Operating and storage ambient humidity:
    10...90 % RH (non-condensing).
    Display: 3 digits + sign
    Analogue inputs: 3 NTC/PTC inputs
    (parameter configurable)
    Digital inputs: 2 voltage-free parameter-
    configurable digital inputs.
    Signal
    Description
    AH3
    High temperature alarm
    AL3
    Low temperature alarm
    The alarm is handled in the same way as
    the other temperature alarms and stan-
    dard signal delays are used.
    DEFROST ALARM
    If defrosting does not reach the end of
    cycle temperature but is terminated by a
    time out, the alarm controller is immedi-
    ately activated.
    An alarm LED remains on and the alarm
    is recorded in the Ad2 label alarm menu.
    Signal
    Description
    Ad2
    End of defrost due to time
    out
    The alarm is automatically reset as soon
    as the next defrost cycle starts. The
    alarm LED can only be manually
    switched off with a silencing procedure.
    EWDR 983-985 /CS (LX)
    3/11
     PAR.
    DESCRIPTION
    RANGE
    DEFAULT
    VALUE
    LEVEL
    U.M.
    SEt
    Set point with range falling between the minimum
    LSE set point and the maximum HSE set point. The
    value of the set point is in the
    machine status
    menu
    LSE...HSE
    0.0
    °C/°F
    diF
    When the compressor stops when it reaches the
    set point value, it restarts at a value corresponding
    to the set point plus the value of the differential It
    must not be 0
    0.1...30.0
    2.0
    1-2
    °C/°F
    HSE
    Maximum set point value
    LSE...302
    50.0
    1-2
    °C/°F
    LSE
    Minimum set point value
    -58.0...HSE
    -50.0
    1-2
    °C/°F
    OSP
    Offset point. Value to be added to set point if
    reduced set point is activated Economy function).
    -30.0...30.0
    0
    2
    °C/°F
    Cit
    Minimum compressor ON time. Minimum time for
    activating a compressor before deactivation Not
    active if=0
    0...250
    0
    2
    min
    CAt
    Maximum compressor ON time. Maximum time for
    activating a compressor before deactivation Not
    active if=0
    0...250
    0
    2
    min
    Ont (1)
    Compressor activation time if probe is faulty. If set
    to 1 with OFt=0 the compressor always remains on
    whereas if OFt>0 it operates in duty cycle mode
    (see
    Duty Cycle
    diagram)
    0...250
    0
    1-2
    min
    OFt (1)
    Compressor shut-down time if probe is faulty. If set
    to 1 with Ont=0 the compressor always remains off
    whereas if OFt>0 it operates in duty cycle mode
    (see
    Duty Cycle
    diagram)
    0...250
    1
    1-2
    min
    dOn
    Delay in activating compressor relay after start-up
    0...250
    0
    1-2
    sec
    dOF
    Delay after shut-down; between compressor relay
    shut-down and subsequent start-up the specified
    time must elapse.
    0...250
    0
    1-2
    min
    dbi
    Delay between switch-ons; the specified time must
    elapse between two subsequent switch-ons
    0...250
    0
    1-2
    min
    OdO
    Delay in activating outputs after start-up of instru-
    ment or after a power failure. Not active if=0
    0...250
    0
    1-2
    min
    dty
    Type of defrost.
    0 = electrical defrosting;
    1 = cycle reversing defrosting (hot gas);
    2 = Free mode defrosting (compressor disabled).
    0/1/2
    0
    1-2
    flag
    dit
    Period of time elapsing between the start of two
    defrosts 0=function disabled
    0...250
    6
    1-2
    hours
    dt1
    Unit of measurement for defrost times (par. dit)
    0=”dit” expressed in hours
    1=”dit” expressed in minutes
    2=”dit” expressed in seconds
    0/1/2
    0
    2
    flag
    dt2
    Unit of measurement for duration of defrosting
    (dEt parameter)
    0= “dEt” parameter expressed in hours
    1= “dEt” parameter expressed in minutes
    2= “dEt” parameter expressed in seconds
    0/1/2
    1
    2
    flag
    dCt
    Selection of defrosting time count mode.
    0=compressor operating hours DIGIFROST®
    method). Defrosting active only if compressor is
    on.
    1 = equipment operating hours; defrost counting is
    always active when the machine is on
    2=compressor stop Each time the compressor stops
    a defrosting cycle according to par. dtY
    3=con RTC. Defrosting at times set by dE1...dE8,
    F1...F8 parameters.
    0/1/2/3
    1
    1-2
    flag
    dOH
    Delay between start of first defrosting operation
    and start-up of instrument.
    0...59
    0
    1-2
    min
    dEt
    Defrosting time-out; determines maximum dura-
    tion of defrosting.
    1...250
    30
    1-2
    min
    dE2
    Defrosting time-out on second evaporator; deter-
    mines maximum duration of defrosting on 2nd
    evaporator
    1...250
    30
    2
    min
    NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the
    level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels
    EWDR 983-985 /CS (LX)
    4/11
    PAR.
    DESCRIPTION
    RANGE
    DEFAULT
    VALUE*
    LEVEL**
    U.M.
    dSt
    End of defrost temperature (determined by evapo-
    rator probe)
    -50.0...150
    8.0
    1-2
    °C/°F
    dS2
    End of defrost temperature (determined from
    probe on 2nd evaporator)
    -50.0...150
    8.0
    2
    °C/°F
    dPO
    Determines when instrument starts up if the
    defrosting cycle must be activated (if the tempera-
    ture on the evaporator allows this) y=defrosting
    activated at start-up n=defrosting not activated at
    start-up
    n/y
    n
    1-2
    flag
    tcd
    minimum time for each compressor state before
    defrosting “Ontime if >0; “Offtime if >0
    -31...31
    0
    2
    min
    Cod
    Compressor “Off” time before defrost cycle. The
    compressor is not turned on if a defrost cycle is
    expected in the time indicated by the parameter.
    0=Function excluded
    0...60
    0
    2
    min
    “dd (2)
    dE1...dE8; daily defrost start time
    0...23/0...59
    24
    1-2
    hours/min
    “Fd (2)
    F1...F8 festive defrost start time
    0...23/0...59
    24
    1-2
    hours/min
    FPt
    Determines if “FSt” and “Fot” are expressed as
    absolute values or in relation to set point
    0=absolute value; 1=value related to set point
    0/1
    0
    2
    flag
    FSt
    Fan stop temperature. Temperature limit that, if
    exceeded by the value read by the evaporator
    probe, stops the fans.
    -50.0...150.0
    2.0
    1-2
    °C/°F
    Fot
    Fan start temperature. If the temperature read by
    the evaporator probe is lower than the set value
    the fans remain off.
    -50.0...150.0
    -50.0
    2
    °C/°F
    FAd
    Fan activation intervention differential. (see “FSt”,
    “Fot”)
    1.0...50.0
    2.0
    1-2
    °C/°F
    Fdt
    Delay before fan activation after defrosting
    0...250
    0
    1-2
    min
    dt
    Dripping time
    min
    0...250
    0
    1-2
    dFd
    Disables evaporator fans. y=fans disabled n=fans
    enabled
    y/n
    y
    1-2
    flag
    FCO
    Disables fans with compressor off (Off)
    y = fans active (with thermostat; in response to
    value read by defrost probe, see “FSt” parameter);
    n = fans off;
    dc = duty cycle (using parameters “Fon” and “FoF”)
    n/y/dc
    y
    1-2
    flag
    Fod
    Enables fan stop with door open and fan re-start
    when door is closed (if fans were on).
    n=fans stop;
    y=fans unchanged
    n/y
    n
    2
    flag
    FdC
    Fan shut-down delay after compressor stop
    0=function excluded
    0...99
    0
    2
    min
    Fon
    Fan start-up time in Duty Cycle mode; valid for
    FCO=dc and H42=1 (presence of probe 2)
    0...99
    0
    2
    min
    FoF
    Fan shut-down time in Duty Cycle mode; valid for
    FCO=dc and H42=1 (presence of probe 2)
    0...99
    0
    2
    min
    Att
    Determines if “LAL” and “HAL” are expressed as
    absolute values or as a differential related to the
    set point
    0=absolute value
    1=value related to set point
    0/1
    0
    2
    flag
    AFd
    Alarm differential
    1.0...50.0
    2.0
    1-2
    °C/°F
    HAL (4)
    Maximum alarm. Temperature limit (whose
    absolute or relative value status is regulated by
    “Att”) above which the alarm is activated.
    LAL...150.0
    50.0
    1-2
    °C/°F
    LAL (4)
    Minimum alarm. Temperature limit (whose
    absolute or relative value status is regulated by
    “Att”) below which the alarm is activated.
    -50.0...HAL
    -50.0
    1-2
    °C/°F
    PAO (5)
    Alarm exclusion time after start-up of instrument
    following a power failure
    0...10
    0
    1-2
    hours
    dAO
    Alarm exclusion time after defrosting
    0...999
    0
    1-2
    min
    NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the
    level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels
    EWDR 983-985 /CS (LX)
    5/11
    [ Pobierz całość w formacie PDF ]
  • zanotowane.pl
  • doc.pisz.pl
  • pdf.pisz.pl
  • jaczytam.opx.pl
  • 
    Wszelkie Prawa Zastrzeżone! Oto smutna prawda: cierpienie uszlachetnia. Design by SZABLONY.maniak.pl.